datasheetdetails
|
<< | < | AX-ANA-LPR - Redox Potential (ORP) Sensor
|
|
The AX-ANA-LPR-xxx is an Analytical sensor for the measurement of Redox potential in liquids. Oxidation or reduction occurs on the platinum surface of the probes The potential difference between the submerged and reference probes generates a voltage proportional to the sum of oxidants and reducers in the measured fluid. There is the option of a relay and / or an LCD Display module. |
Product Features
| • Measuring Range +/- 1000mV • Easy to set-up
| • Mounted on the top of a tank • Variety of probes available
|
Measuring Range
| +/- 1000mV Reserve +/-200mV
| Resolution
| 0.1mV (internal resolution 0.8mV)
| Linearity
| +/- 0.001%
| Accuracy
| 0.1% of the measured value +/- 1 digit +/- 0.01% /Deg C
| Measuring Rate
| 300msec, on the display (refreshing rate) 1 sec
| Process Temperature
| -50 to +100 degC (semi-conductive sensor)
| Power Supply
| 12 to 36Vdc 720mW galvanically isolated, protected against surge transients
| Probes - LPR-111-X
| 50 uS/cm 6 bar 80 C with solid particles
| Probes - LPR-121-X
| 500 uS/cm 6bar@25C 100C with solid particles
| Probes - LPR-131-X
| 150 uS/cm 0.5bar 60C clear liquid
| Probes - LPR-141-X
| 150 uS/cm 3bar 60C clear liquid
| Probes - LPR-151-X
| 150 uS/cm 6bar 80C clear liquid
| Probes - LPR-161-X
| 150 uS/cm 3bar 60C clear liquid horizontally mountable
| Output - LPR-1X1-4
| HART & 4-20mA
| Output - LPR-1X1-8
| HART & 4-20mA EeX
| Output - LPR-1X1-H
| HART & 4-20mA & 1 x SPDT 1A@30Vdc
| Display (option) SAP300
| Full featured programming LCD Display
| Intrinsically Safe Class
| ATEX II 1 G Exia B T6 Ga
| Cable Length
| 5m provided
| Weatherproof Rating
| IP68
| Materials Housing
| Plastic
| Process Connection
| 1 1/2" BSP Polyproylene as standard ( PVDF option at additional cost) NPT option
| Electromagnetic Compatability
| EN61326 Class B
| Country of Origin
| EU
|
AX-ANA-LPR.pdf / Analytical sensor -Redox potential |  |
|
|
|